Fluoroscopic Capacity
|
Max rated capacity
|
3.5 kW
|
Max rated fluoroscopic capacity
|
Tube Current 4mA; Tube Voltage 110kV
|
Automatic Fluoroscopy
|
Tube Voltage: 40kV~110kV adjust automatically Tube Current: 0.3mA~4mA adjust automatically
|
Manual Fluoroscopy
|
Continuous tube Voltage: 40kV~110kV Continuous tube Current: 0.3mA~4mA
|
Pulse Fluoroscopy
|
Continuous tube Voltage: 40kV~110kV Continuous tube Current: 4.1mA~8mA
|
Tube Voltage, mAs
|
40kV~110kV 20~63mA 1~125mAs
|
Plateholde size
|
200mm×250mm(8"×10") or 250mm×300mm(10"×12")
|
X-ray Tube
|
X-ray tube special for High Frequency
|
Large focus: 1.5, small focus: 0.6 Inverter Frequency: 40kHz Thermal capacity: 30kJ (40kHU)
|
Video System
|
Image Intensifier
|
Image Intensifier made by TOSHIBA (9")
|
CCD vidicon
|
Imported CCD Vidicon with ultra-low luminosity
|
Monitor
|
Horizontal 1000 lines and vertical 800 lines, Bandwidth: 12.5MHz, Image/sec: 25
|
CCU (central control)
|
Recursive filter: K=8, 7 images storage, image upright, image overturn, positive & negative image; LIH(last image freeze, and OSD(monitor display)
|
Structure
|
Direction-wheel
|
±90°revolution, can change the moving direction of the unit.
|
Ascending & descending range of pillar
|
≥400mm
|
C-arm
|
Forward and backward movement: 200mm Revolution around horizontal axis: ±180° Revolution around vertical axis: ±15 Slip on orbit: 120°(+90°~ -30°)
|