X-ray Generator
|
Rotary anode, double focus, and bridge silicon rectification. The X-ray generator is an oil-immersed self-cooled unit.
|
Max rated capacity
|
|
Current 15mA 30mA 60mA 100mA Voltage 90kVp 90kVp 90kVp 80kVp Time 6.3s 6.3s 4.0s 3.2s
|
Power supply
|
Voltage: 180~240V; Frequency: 50Hz Power: not lower than 7kVA; Resistance: not larger than 1Ω
|
Voltage adjustment range of X-ray tube
|
45~90kVp, continuous adjustment
|
Range of the timer
|
0.08~6.3s
|
Fluorescent screen size
|
356 mm × 356 mm
|
Distance between the focus to the screen
|
max: 830 mm; min: 630 mm
|
Tube head carriage movement range
|
vertical: 1100 mm; horizontal: 260 mm
|
Beam limiter
|
field coverage at 100 cm FFD, 45×45 cm (max); 0×0 cm (min)
|
Specification of X-ray tube
|
Model XD55 rotation anode, double focus 1.8/4.3mm
|
Movement of the tabletop
|
400mm
|
Field size of spot film device
|
127mm×178mm (1 pc); 203mm×254mm (1 pc); 1/2 (127mm×178mm) (2 pcs)
|
Transport dimension (L×W×H) (mm)
|
2310×1000×1250
|
Weight (kg)
|
Net: 418 / Gross: 564
|