1. Single bed with single tube
|
2. Design according to the electric mechanism of middle-size machine, with complete and reasonable function.
|
3. High-voltage system is of split structure, with high-voltage generator of full-wave rectification, independent X-ray tube unit, large output power and reliable and stable operation.
|
3. Power voltage(V),photograph kilovolt(kV) and infinitely variable control.
|
4. Be equipped with the manostat for the filament of X-ray tube and space charge complementor to ensure the stable and exact output.
|
5. Be equipped with capacity limit potective circuit ---kV mA s interlock limit protection.
|
6. The timer circuit grades according to R10 priority comfficient has an exact time control.
|
7. The photographic flat bed has a vibrating ray-filter.
|
8. The pillar has only the bottom track, with easy installation.
|
Pillar
|
Move Lengthwise
|
1500mm
|
Move Transverse
|
360mm
|
Move Up and Down
|
720~1720mm (from focus to ground)
|
Power Supply
|
Single phase 380V/220V, 50Hz±0.5Hz, 10kVA, 380V : 2Ω; 220V : 0.6Ω
|
|
Max rated capacity
|
|
Current 25mA 50mA 100mA 125mA Voltage 90kV 90kV 85kV 82kV Time 1.0s 6.3s 2.5s 1.0s
|
Max DC output voltage
|
|
60KV
|
Max DC output current
|
|
100mA
|
Fluoroscopy
|
Voltage
|
50~90KV
|
Current
|
0.5~5mA
|
Photography
|
Voltage
|
50~90KV
|
Current
|
Small focus 15mA; big focus 50, 60, 100mA 4 grades together
|
Spot film
|
100mA fixed
|
Time
|
0.04~6.3s; 24 grades together
|
Generator
|
Capacity
|
100KVA
|
X-ray Tube
|
XD4~2.9/100; double focus 1.8/4.3
|
Table
|
Rotary Angle
|
+90°~0°~-5°
|
Film Size
|
128×179mm (5"×7"); 205×254mm (8"×10"); 254×305mm (10"×12")
|